Measuring ranges | Smallest measuring range: 0…30 mbar |
| Largest measuring range: 0…40 bar |
| Negative pressure measuring range: up to -1 bar |
Analog output | 4…20 mA, 2-wires |
| 4…20 mA, 2-wires, II 1G Ex ia IIC T4 Ga |
| 0…10 V, 3-wires |
Power supply | 20 mA output: 9…30 V DC |
| 10 V output: 15…30 V DC |
Accuracy | ≤ ± 0,2 % FS @ 25 °C |
| ≤ ± 0,5 % FS @ 25°C ≤ 60 mbar |
Long term stability | ≤ ± 0,15 % FS / Year |
Response time | 200 ms - others on request |
Switch-on time | < 1 s |
Temperature coefficient | Zero: ≤ ± 0,015 % FS / Kelvin |
| Span: ≤ ± 0,01 % FS / Kelvin |
Medium temperature | -40…100 °C (125 °C < 0,5 h) |
| -20…80 °C with EX ia IIC T4 |
Surrounding temperature | -25…80 °C |
| -20…80 °C with EX ia IIC T4 |
Storage temperature | -40…85 °C |
Short-circuit resistance | Permanent |
Reverse polarity protection | Protection against reverse polarity, but no function |
Electromagnetic compatibility | Interference emissions and immunity acc. to EN 61326 |
Vibration | 4g with 3-axis resonance frequency according to DIN EN 60068-2-6:2008 |
Wetted materials | Process connection: Stainless steel 1.4404 |
| Sensor: Ceramic Al2O3 - FDA |
| Process seal: FPM (Viton), NBR, EPDM, FFKM (Chemraz / Kalrez) |
Protection type | IP 65 |
| IP 67 with cable connection |
Weight | Approx. 250 g |